Pre-treatment 2D and 3D dosimetric verification of volumetric arc therapy. A correlation study between gamma index passing rate and clinical dose volume histogram

0

by Lukasz Szczurek, Robert Juszkat, Jolanta Szczurek, Ilona Turek, Piotr Sosnowski

Objectives

To evaluate methods for the pre-treatment verification of volumetric modulated arc therapy (VMAT) based on the percentage gamma passing rate (%GP) and its correlation and sensitivity with percentage dosimetric errors (%DE).

Methods

A total of 25 patients with prostate cancer and 15 with endometrial cancer were analysed. The %GP values of 2D and 3D verifications with different acceptance criteria (1%/1 mm, 2%/2 mm, and 3%/3 mm) were obtained using OmniPro and Compass. The %DE was calculated using a planned dose volume histogram (DVH) created in Monaco’s treatment planning system (TPS), which relates radiation dose to tissue and the patient’s predicted dose volume histogram in Compass. Statistical correlation between %GP and %DE was verified using Pearson’s correlation coefficient. Sensitivity was calculated based on the receiver operating characteristics (ROC) curve. Plans were calculated using Collapsed Cone Convolution and the Monte Carlo algorithm.

Results

The t-test results of the planned and estimated DVH showed that the mean values were comparable (P > 0.05). For the 3%/3 mm criterion, the average %GP was acceptable for the prostate and endometrial cancer groups, with average rates of 99.68 ± 0.49% and 99.03 ± 0.59% for 2D and 99.86 ± 0.39% and 99.53 ± 0.44% for 3D, respectively. The number of correlations was poor for all analysed data. The mean Pearson’s R-values for prostate and endometrial cancer were < 0.45 and < 0.43, respectively. The area under the ROC curve for the prostate and endometrial cancer groups, was lower than 0.667.

Conclusions

Analysis of the %GP versus %DE values revealed only weak correlations between 2D and 3D verifications. DVH results obtained using the Compass system will be helpful in confirming that the analysed plans respect dosimetric constraints.