Добавить новость
ru24.net
News in English
Март
2022

Evaluation of ocular biometry in primary angle-closure disease with two swept source optical coherence tomography devices

0

by Boonsong Wanichwecharungruang, Anyarak Amornpetchsathaporn, Wisakorn Wongwijitsook, Kittipong Kongsomboon, Somporn Chantra

Purpose

To investigate agreement between 2 swept source OCT biometers, IOL Master700 and Anterion, in various ocular biometry and intraocular lens (IOL) calculations of primary angle-closure disease (PACD).

Setting

Rajavithi Hospital, Bangkok, Thailand.

Design

Prospective comparative study.

Methods

This study conducted in a tertiary eye care center involving biometric measurements obtained with 2 devices in phakic eye with diagnosis of PACD. Mean difference and intraclass correlation coefficient (ICC) with confidence limits were assessed, and calculations of estimated residual refraction of the IOL were analysed using Barrett’s formula.

Results

Sixty-nine eyes from 45 PACD patients were enrolled for the study. Excellent agreement of various parameters was revealed, with ICC (confidence limits) of K1 = 0.953 (0.861–0.979), K2 = 0.950 (0.778–0.98), ACD = 0.932 (0.529–0.978), WTW = 0.775 (0.477–0.888), and LT = 0.947 (0.905–0.97). Mean difference of axial length (AL) was -0.01 ± 0.02 mm with ICC of 1.000. IOL calculation was assessed with Barrett’s formula, and Bland-Altman plot showed excellent agreement in the results of the 2 devices for the IOL power and estimated post-operative residual refraction (EPR).

Conclusions

Mean differences of biometric parameters, obtained with IOL Master700 and Anterion, were small, and ICC showed excellent concordance. No clinical relevance in calculation of IOL power was found, and the two devices appeared to be comparably effective in clinical practice.




Moscow.media
Частные объявления сегодня





Rss.plus
















Музыкальные новости




























Спорт в России и мире

Новости спорта


Новости тенниса